Practical Example: Counting Small Parts
Menu code settings used in the example:
Function Code
Counting 2 1 4
Ref. qty. and ref. weight 7 1 2
With data ID code 7 2 2
Application: Counting bulk quantities of items with the same weight
Step/Key Readout Data Output/Printout
Place container on pan +
22.65 gcs
c, t
0.00 gcs
Add 10 counted parts
Press v to confirm reference rEF
10 pcs
nRef + 10 pcs
sample quantity +
5.6546 gcs
wRef + 5.6546 g
+
10 pcs
Fill container with desired
quantity of parts
(without counting them) +
500 pcs
p +
500 pcs
Qnt + 500 pcs
Press c to clear the
weight readout and
the reference value +
2827.35 gcs
p +
2827.35 gcs
N+2827.35g
The data ID codes mean:
nRef + 10 pcs reference sample quantity (piece count)
wRef + 5.6546 g reference weight
Qnt + 500 pcs calculated piece count
N+2827.35gnet weight