Deleting Calibration Data
Intended Use
Example: Here, delete calibration of the pH measurement
pH measurement: measuring at default settings
Redox potential: Delete the mV offset and return to the absolute mV mode.
Step Press key Display
1. Activate menu m
2. Confirm calibration Press q softkey;
functions and then pH if necessary press
O softkey repeatedly
3. Select “Clear”
q softkey,
and confirm
O softkey
4. Delete: select “Yes” and
q softkey,
confirm
OK softkey
19
Cal. Functions
Cal. Record
Erase Cal
Select Buffer Set
Erase Cal.
Yes
>No