Operating the Meter: Measurement
Preparation
§ To select measuring mode: Press softkey
pH or mV.
!Improperly standardized electrodes
yield inaccurate measurements.
Perform calibration regularly prior to
measurement.
$ Temper buffer or buffer solutions or
measure current temperature, if the
measurement is performed without an
ATC probe.
ATC Probe
Measurements can be performed with
and without an ATC probe.
The Docu-pH
Meter
detects the following
type of ATC probes automatically:
– NCT10 kohm: +14°C to 76°C
– NCT30 kohm: –5°C to 40°C
– PT1000: 4°C to 105°C
!It is imperative that you measure the
temperature to achieve accurate pH
measurements. When measuring with-
out an ATC probe, enter the current
temperature manually:
1) Determine the current temperature
using a thermometer.
2) Start temperature entry:
Softkey
Temp
3) Set temperature value:
Softkey
q, softkey Q
3) Save temperature value:
Softkey
OK
Lock Function
The lock function freezes the value
in the display. Once the stability
conditions have been met, the value
is displayed statically (Symbol: “ “)
Deactivate lock function:
Press Softkey
Unlock
Activate function: see chapter entitled
“Default Settings.”
Fast Mode (Shorter Measurement
Time)
The fast mode reduces the measuring
time by using a lower stability criterion.
Activate function: see chapter entitled
“Default Settings.”