EasyManua.ls Logo

Sartorius Docu-pH - Electrode with DocuClip; DocuClip Intended Use; DocuClip Usage Guide

Sartorius Docu-pH
38 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Electrode with DocuClip
®
Intended Use
DocuClip
®
performs the following functions:
The Sartorius “Docu-pH
Meter
” automatically recog-
nizes any connected electrode and stores the current
calibration value from the DocuClip
®
.
Permanently stores the serial number and initial
calibration data of an electrode.
Displays 4 calibration data records via Docu-pH
Meter
.
Sartorius PY-P...DOC electrode models:
Autom. stores the serial number of the connected
electrode
Product name
Type of electrode
Calibration data logging
Usage
DocuClip
®
communicates with the Sartorius
Docu-pH
Meter
.
DocuClip
®
can be used with just about all electro-
chemical sensors, such as all pH electrodes that can
be connected to the Docu-pH
Meter
.
DocuClip
®
is irreversibly fastened to the cable of the
sensor.
DocuClip
®
is preformatted; easy-to-understand
user guidance prompts in the Docu-pH
+
enable the
user to intuitively perform initialization and initial
calibration.
Using Sartorius electrodes with DocuClip
®
(PY-P...DOC):
The factory-set initial calibration is displayed as the
last of the calibration values and cannot be over-
written.
Using DocuClip
®
with any type of electrode:
Installing DocuClip
®
: If no calibration data are avail-
able, Docu-pH+ prompts the user to manually enter
an electrode number and the initial calibration data.
The initial calibration value will be displayed as the
last of the calibration values and cannot be
overwritten.
26