Docu-pH
+
Meter
Model: Data Output
Print Data Log
Intended use
For printing out measurements, calibration data and ID codes in accordance with the
requirements of quality management systems.
Preparation:
! Switch off numerator on the optional Sartorius printer.
The data log in the pH mode can include the following lines:
. . . . . . . . . . .
. Dotted line
Blank line
No: 4 Stable Consecutive number and measurement status
pH: 6.63 pH
mV: If required, mV in the mV mode
´C: 25.0 MTC Temperature is set manually here
D/T: 20.05.05 11:53 Date and time
ID: 135933 Sample ID, entered by user
CAL: 19.05.05 08:52 Date and time of the calibration
(forms the basis for calculating the pH)
The data log for calibration results:
. . . . . . . . . . . . Dotted line
Blank line
Cal. Record
Sartorius Docu-pH+Meter Instrument number
S/N: 1234567800 Docu-pH
Meter
serial number
pH Electrode PY-P12DOC Type of electrode
S/N: A044805002 Serial number of the electrode
20.05.05 11:56 Date and time of calibration end
S1: 97.2 % Slope 1 (between buffer 1 and buffer 2)
S2: 97.1 % Slope 2 (between buffer 2 and buffer 3)
Offset : -9.1 mV Zero point in mV
(= deviation ... at pH = 7.00)
B1: 1.68 Nominal value of buffer 1 at 25°C
+310.8 mV +25.0 ´C Voltage and temperature of buffer 1
B2: 6.86 Nominal value of buffer 2 at 25°C
+0.1 mV +25.0 ´C Voltage and temperature of buffer 2
Calibration OK Calibration status
Data log from memory identifiable in line:
ID: 135933 memory Sample ID and memory printout
30