The following data are stored permanently in DocuClip
®
:
– Initial calibration including date and time
– pH and voltage in mV
– Nominal values of the buffers used
– Serial number of Docu-pH
Meter
– Serial number of the electrode
– Temperature
§ View data of installed electrode/DocuClip
®
:
See Setup-Menu: Calibration functions: pH/mV: Data DocuClip
®
Enter serial number of the Electrode on Docu-pH
+
:
Step Press key Display
1. Connect DocuClip
®
2. Start “Serial number input” Start softkey
3. Set “Electrode number” (S/N) Repeatedly
press Q softkey,
O softkey
4. Confirm “Electrode number”
OK softkey
!Initial calibration: Please work
with particular care. The initial
calibration documents the degree
of effectiveness of the new electrode.
This calibration is the reference point
for the later analysis whenever there
is a change in performance.
27