Calibration and Adjustment
Purpose
Calibration is the determination of any
difference between the measured value
displayed and the true weight (mass) of
a sample. Adjustment is the correction
of this difference, or its reduction to
an allowable level within maximum
permissible error limits.
Features
Calibration | adjustment can be
performed only when
there is no load on the balance
the balance is tared
the internal signal is stable
the weight displayed for the sample on
the balance must not differ from the
nominal weight by more than 2%.
If these conditions are not met, an error
message is displayed Err 02".
Following calibration | adjustment,
the application program is cleared.
Internal Calibration | Adjustment
Note:
Only for models with the label
Entris...i-1S !
In the operating menu, select
Cal.Just. - cal.int. before beginning.
The built-in motorized calibration
weight located in the housing is applied
and removed automatically for internal
calibration.
t Select calibration | adjustment:
Press Q
> The internal calibration weight is
applied automatically
> The balance is adjusted | calibrated
> The built-in calibration weight is
removed