Error codes are shown on the main display for approx. 2 seconds.  
The program then returns automatically to the previous mode.
Display Cause  Solution
No segments appear on   No AC power is available  Check the AC power supply 
the display  The power supply is not  Plug in the power supply 
  plugged in 
high  The load exceeds the  Unload the balance 
  balance capacity
low or Err 54  Something is touching the  Move the object that is touching 
  weighing pan  the weighing pan
App.err.  Cannot store data:  Increase load 
  Load on weighing pan too  
  light or no sample on pan  
  while application is active
dis.err.  Display error:  Change the configuration in the  
  Data output not compatible   operating menu 
  with output format 
prt.err.  Interface port for printer  Reset the menu factory settings 
  output is blocked  or 
     Contact your local  
Sartorius Service Center 
err  02  Calibration parameter not met,   Calibrate only when zero is displayed 
 e.g.:  
  – Press w to tare the balance 
  – load on weighing pan  Unload the balance
err  10 The w key is blocked  After the tare memory has been 
  for active application programs;   deleted using the c key,  
  Only 1 tare function can be   the w key can be used again 
  used at a time
err  11  Tara memory not allowed   Press w 
The weight readout   Unstable ambient conditions  Set up balance in another area 
changes constantly  (excessive vibration or draft)    Adjust Setup configuration 
  A foreign object is caught  Remove the foreign object 
  between weighing pan   
  and housing
The weight readout  The balance was not  Calibrated/adjust the balance   
is obviously wrong  calibrated/adjusted 
  Balance not tared before  
 weighing
If any other errors occur, contact your local Sartorius Service Center.
Web address: http://www.sartorius.com
Troubleshooting Guide