Preparation
t Select the Counting application in the 
menu:  
see “Configuration"
t Set the following parameters:
Applic.ation programs
  Count. 
  Resolut.ion
    o Disp.Acc. Display  accuracy 
      10-fold  10-fold higher 
  Ref.Updt.  Autom. ref. sample 
updating
      o  Off Off 
      autom.  Automatic
  o = Factory setting
Printout: Counting
 n R e f        1 0   : 
Ref. sample quantity
 wRef    21.14 g  :  Reference weight  
for 1 unit
 Qnt   +   500 pcs  :  Calculated quantity
23
Application Programs
Counting
Display symbol: Z
Purpose
With the Counting application, you can 
determine the number of parts that each 
have approximately equal weight. To 
do this, a known number of parts (the 
reference sample quantity) is weighed first, 
and the individual piece weight (reference 
weight) is calculated from this result. Thus 
the number of parts subsequently placed 
on the balance can be determined from 
their weight.
Changing the Reference Sample Quantity
Activate function:  
Press the b key  
Select the desired reference sample 
quantity (1 to 100): 
In increments of 1: Press the b key briefly 
In increments of 10:  
Press and hold the b key.
The quantity is stored in battery-backed 
 
memory.
Reference Sample Updating
Automatic reference sample updating 
optimizes the counting accuracy. You can 
activate or deactivate this function in the 
menu.
Automatic reference sample updating 
is performed when the requirements, 
including the specified stability criterion, 
have been met. 
The abbreviation opt for “optimizing”, is 
displayed briefly with the new reference 
sample quantity.