28
Example: Determining animal weight with automatic start and 20 subweighing operations
(measurements)
Parameter settings: Applic. - Animalw. (menu code 2. 7.)
Step Key (or instruction) Display/Data output
1. Place animal weighing
22.6 g
bowl on the balance
2. Tare the balance w 0.0 g
3. Change the number of
b ref 30
subweighing operations:
4. Select number of measurements: Repeatedly:
b
In increments of 1 (1, 2, 3, ..., 100) Press briefly
b
In increments of 10 (10, 20, ..., 100) press and hold
ref 20
5. Confirm number of measurements
u + 0.0 g
*
and start automatic animal weighing.
The number of measurements remains
stored in battery-backed memory
until the setting is changed.
6. Place first animal in bowl.
888
The balance delays the start
19
of measurements until the
20
difference between 2 measure-
....
ments meets the criterion.
1
7. Read off the result. + 410.1 g
G
*
The result is displayed with
the “*” symbol (= calculated value)
m D e f 2 0
and remains displayed until the
x-Net + 410.1 g
sample (animal) is removed
from the load plate (bowl).
8. Unload the balance +
0.0 g
*
9. Weigh next animal (if des.)
Next weighing series begins automatically.