The following functions are available
in the SETUP menu, under the isoTEST
menu item:
Weighing system:
Calibration/adjustment
Weighing only
Hardware tests
Test interfaces
Heater test
When the sample pan and the pan sup-
port are removed from the sample
chamber, you have access to the follow-
ing functions:
Heater adjustment
2-point temperature adjustment
1-point temperature adjustment
31
“isoTEST” Calibration/Adjustment Functions
Heater Adjustment
Using 1-point and 2-point temperature
adjustment and the YTM03MA temper-
ature adjustment set (see “Accessories”)
you can calibrate and adjust temper-
ature settings of the drying unit.
Weighing System Settings
Calibration, Adjustment
Purpose
Calibration is the determination of the
difference between the weight readout
and the true weight (mass) of a sample.
Calibration does not entail making any
changes within the weighing system.
Adjustment is the correction of the
difference between the measured value
displayed and the true weight (mass)
of a sample, or the reduction of the
difference to an allowable level within
maximum permissible error limits.
Features
Calibration is performed externally
with a pre-set weight value of 30 g (see
“Accessories”, order no. YSS43).
You can have calibration and adjust-
ment results documented as a ISO/GLP-
compliant printout; see the next page.