Step Key (or Instruction) Display/Printout
--------------------
13.08.2001 16:01
Model MA45-000230V
Ser. no. 13206969
Ver. no. 01-43-01
ID
--------------------
Prg. 1
Heating STANDARD
Fin. temp. 130 ’C
Stdby temp. OFF
Start W/STABIL.
End AUTOMATIC
IniWt + 2.036 g
The current moisture loss is displayed
after the data is printed.
The drying program shuts off
automatically once no further
weight loss is detected (in this
example, after 5.2 minutes)
Then the footer of the analysis --------------------
record is printed FinWt + 1.814 g
5.2+ 18.90 %L
Name:
--------------------
30
005
PROGRAMM
P1 130
o
C Auto. 37
o
C 0.8min
S
%L
CANCELSETUP
1890
PROGRAMM
P1 130
o
C Auto. 128
o
C 5.2min
S
%L
NEXTSETUP