Getting Started
  Cubis MSE Operating Instructions  19
Leveling the Balance, Setting the Level Indicator
h
  Leveling the balance compensates for slant or unevenness at the place 
of installation� The balance must be perfectly horizontal to ensure 
consistent, reproducible weighing results� All models are equipped with 
an electronic tilt angle detection feature� If the balance is not level,  
a warning symbol   is shown on the display� Push-button automatic 
leveling is available on models with motorized leveling feet� For models 
with manual leveling feet, please follow the instructions on the display: 
see “Leveling the Balance with the Inclination Sensor�" 
Leveling the Balance Manually
t  The balance is leveled using both front leveling feet�
t  Screw in both back leveling feet (only for models with back leveling feet)� 
t  Turn the two front feet as shown in the illustration until the air bubble is 
centered within the circle of the level indicator�
y  Normally, several leveling steps are required�
t  Screw out both back leveling feet until they touch the setup surface  
(only for models with back leveling feet)�
Warm-up Time
y  In order to deliver the precise results described in the Technical Specifications, 
the balance must warm up for at least 30 minutes after initial connection to 
the power supply� Only after this time will the device have reached the required 
operating temperature�
  When a verified balance used in legal metrology (legal-for-trade 
applications) is connected to the power, it must warm up for at least 
two hours before operation�