Cubis MSE Operating Instructions  9
Getting Started
Getting Started
Unpacking the Equipment
t  Open the packaging at the top�
t  Use both hands to lift the balance, with the packaging, out of the cardboard 
box�
t  Place the packaging with the balance on the floor�
t  Remove the top part of the packaging�
Installation Instructions
t  Select the proper setup location:
–  Set up the device on a stable, even surface that is not exposed to vibrations 
(e�g� weighing stone)�
–  Place the device in a location with enough free space around it so that 
excessive heat cannot build up� 
–  Maintain free access to the device at all times�
  Choose a location that is not subject to the following negative influences:
–  Heat (heater or direct sunlight)
–  Drafts from open windows, AC systems and doors
–  Extreme vibrations during weighing  
–  Excessive moisture
Acclimatization 
Condensation from humidity can form on the surfaces of a cold device when it is 
brought into a substantially warmer area�  
To avoid the effects of condensation, condition the balance for about two hours at 
room temperature, leaving it unplugged from the power supply�
Carefully read all warnings and safety precautions in the respective section of this 
manual�