26 Cubis MSE Operating Instructions
Getting Started
t Attach the display and control unit to the electronics module:
Reattach the two retaining screws�
Open the slot for the connection cable on the electronics module:
t Remove the screws from beneath the electronics module and detach the plate�
Connect the display and control unit to the electronics module:
t Plug in the connection cable�
t Then reattach the plate to the slot�
t Slide the cable that sticks out into the cable channel�
Preparing Below-Cell Weighing
Balances with a readability of 0,01 mg:
A port for a below-cell weighing hook is located on the bottom of the balance�
Analysis and Precision Balances with a Weighing Range up to 15 kg:
1. Remove the hook for below-cell weighing from the clip on the bottom of the
balance�
1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius MSE20201S and is the answer not in the manual?

Sartorius MSE20201S Specifications

General IconGeneral
Readability0.1 g
Repeatability (Std. Dev.)0.1 g
Pan Size400 mm x 300 mm
Power SupplyAC Adapter
DisplayLCD
Stabilization Time (typical)1.5 s
Operating Temperature+10°C to +30°C
Linearity0.2 g
CalibrationExternal
InterfaceRS-232