Error Messages
Error Messages
Error codes are displayed for about 2 seconds� The program then returns automatically to the weighing mode�
Display Cause Solution
No segments appear No power 
Power supply not plugged in
Check power supply 
Connect power supply to the wall outlet (mains)
high
Weighing capacity exceeded Unload the weighing pan
low or Err 54 Contact between load plate and environment Move the object that is touching the weighing pan
App.err.
Cannot save data: load is too light or  
no sample on the weighing pan while  
application is active
Increase load
dis.err.
Data output not compatible with output format Change the configuration with output format in 
the operating menu
prt.err.
Interface port for printer output is blocked Reset the menu factory settings or  
Contact your local Sartorius Service Center
modul.err.
Balance cannot find a Bluetooth printer  
No Bluetooth connection
Establishing a radio connection between the bal-
ance and printer: 
See “Establishing a Radio Connection with the 
Bluetooth
®
 Printer"
err  02
Adjustment condition was not met, e�g�,  
– Tare with the J key 
– Load on weighing pan 
Calibrate only when zero is displayed 
 
Unload the balance
err  10
The J key is blocked when the “Net-total 
Formulation" application is active; only 1 tare 
function can be used at a time
The J key can be used again after the tare 
memory has been deleted using the F key
err  11
Tare memory not allowed
Press the J key
err  81
1)  No print command received from the bal-
ance� Bluetooth
®
 connection was briefly 
interrupted�
2)  Balance was turned on and then data output 
took place�
Wait approx� 40 seconds; device will reconnect 
automatically� 
 
Reestablish connection: See “Establishing a Radio 
Connection with the Bluetooth
®
 Printer"
level.err.
You should level the balance Level balance
 flashes
Balance not level 1)  Level the balance using the leveling feet
2)  or for models with motorized leveling feet:  
Press the * key
The weight readout 
changes constantly
Setup location unstable (excessive vibration or 
draft) 
Foreign object is caught between weighing pan 
and balance/scale housing
Set up balance in another area  
Change Setup Configurations 
Remove the foreign object
The weight readout is 
obviously wrong
The balance was not calibrated/adjusted  
Balance was not tared before weighing
Adjustment 
Tare
If any other errors occur, contact your local Sartorius Service Center.
Contact information: Please point your Internet browser to: http://www�sartorius�com