Operation
30  Cubis MSE Operating Instructions
Basic Weighing Function
 Features
–  Tare the balance
–  Print weight value
  Using a verified balance in legal metrology: 
The Type-Approval Certificate for verified balances is only valid for non-automatic 
weighing instruments�  
For automatic operation with or without additional, integrated equipment, please 
follow the applicable national regulations for the installation location� 
§   Before use in legal metrology, the balance should be adjusted via the internal 
motorized calibration weight at the installation location:  
see the “Calibration and Adjustment" section in this chapter�
$ 
The temperature range (°C) listed on the ID label should not be exceeded during 
operation�
 Example:  
 MSX 
  K 
 +10/+30°C 
  +5°C/+40°C isoCAL
 Preparation
§  Turn balance on: Press the A key 
§  Level the balance if required 
§  Tare the balance if necessary: Press the J key
$  If necessary, change the configuration settings: see “Configuration"
$  If desired, load the factory settings: see “Configuration"
  Additional functions:
$  Turn balance off: Press the A key
  Quick Guide: First Weighing
  Step  Press key  Display/Printout
1�  Turn balance on  A       0.0 g
   Self-test  runs�   
Then the balance performs an automatic initial tare�
2�  Place container on weighing pan (in this example : 11�5 g)�    +         11.5 g
  Close the draft shield (if available)�
3�  Tare the balance  J       0.0 g 
  The balance is tared and the displayed value is zero�
4�  If required, open the draft shield and carefully place the sample in 
  the container on the weighing pan - in this example : 132 g�    +  132.0 g
  Close the draft shield (if available)� You can read the measured value   
   as soon as the weight value stops changing and the unit is displayed�
5�  Print weight value  P  N     +    132.0 g