Care and Maintenance
80  Cubis MSE Operating Instructions
Transporting the Balance
If repairs are required, use the original packaging to transport the balance� 
Toensure adequate protection for safe shipment, Sartorius products have been 
packaged to the extent necessary using environmentally friendly materials� Only the 
original packaging provides optimum protection for the equipment�
    t  Disconnect the device from the power supply�
t  Disconnect any data cables from the device�
  t  Remove all items (such as weights, sensors, etc�) from the weighing chamber�
1.  Remove the other side draft shield panel 
2.  Remove the other side draft shield panel 
3.  Remove the upper draft shield panel 
4.  Remove the weighing pan
5.  Remove pan support (not for MSE225…/MSE125… models)
6.  Remove shield plate/draft shieldz
  t   On models without a draft shield: press down on the two pan support 
fasteners�
t  Rotate and remove the pan support�
  t  Place the balance in the lower part of the packaging�
2
1
3
4
5
6