Specifications
Model R 160 P
Weighing capacity g 30/60/162
Readability mg 0.01/0.02/0.05
Tare range (by subtraction) g -162
Standard deviation mg
±0.02/0.03/0.05
Max. linearity mg
±0.03/0.05/0.1
Stabilization time (typical) s 5
Display update s 0.2
Adaption to ambient conditions
and application requirements
by selection of one
of 4 optimized filter levels
Stability range d 0.25 ... 64 (selectable)
Ambient temperature range K 283-313 (+10
o
C ...+40
o
C)
Sensitivity drift within 283 … 303 K /K
± 1 x 10
–6
Pan size mm
90
Balance housing (W x D x H) mm 219 x 408 x 317
(8.6 x 16.1 x 12.5 in)
Weighing chamber (W x D x H) mm 188 x 149 x 253
7.4 x 5.9 x 10 in)
Net weight kg 13 (27.5 lbs)
Power requirements
(voltage + frequency)
100 / 120 / 220 / 240 V,
50-60 Hz
Allowable voltage fluctuation -15%…+10%
Power consumption VA 12
Interface RS 232C / V24-v28
RS 423 V 10; 7 bit
parity: even, odd, mark, space;
transmission rates: 150… 9600 Baud