EasyManua.ls Logo

Sartorius Sartocheck 3 plus - Calibration and Adjustment Results

Sartorius Sartocheck 3 plus
107 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
44
During calibration, the results are displayed as a deviation from the adjustment
curves. If the deviation is below 0.1 % FS, an adjustment is not required.
At deviations of 0.1-0.2 % FS, an adjustment is recommended. At deviations
> 0.2 % FS, an adjustment is mandatory.
During adjustment, a new adjustment curve is calculated and activated using
the entered values. The data of the adjustment can be printed out on the internal
printer after the adjustment is completed. The new adjustment curve can then
be checked with a calibration.
Fig. 7-67 | Result of the adjustment
Fig. 7-66 | Result of the calibration

Table of Contents

Related product manuals