Preparation
§ Switch on the scale: Press e
§ While all segments are lit, 
press the ) key
§ Select the Application menu: Press k
repeatedly until 
Appl is displayed
§ Open the Application menu: 
Press the ) key
§ Select the Counting application:
Press the k key repeatedly until the
desired menu item is displayed and
press ) to open the submenu
Application Parameters: Counting
3.6. Minimum Load for Initialization
3.6.1* 1 digit
3.6.2 2 digits
3.6.3 5 digits
3.6.4 10 digits
3.6.5 20 digits
3.6.6 50 digits
3.6.7 100 digits
3.6.8 200 digits
3.6.9 500 digits
3.6.10 1000 digits
3.9. Resolution for Calculation of 
Reference Value
3.9.1* Display resolution
3.9.2 Display resolution +
1decimal place
3.9.3  Display resolution +
2decimal places
3.9.4 Internal resolution
3.11 Storage Parameter
3.11.1* At stability
3.11.2 At increased stability
3.12. Reference sample updating 
3.12.1 Off
3.12.3* Automatic
3.13. Reference Weighing Instrument
3.13.1* No reference instru-
ment selected
3.13.2 WP1
3.13.3 WP2
* = Factory setting
§ Press ) to save your settings and (
(repeatedly) to exit the operating menu. 
Storage Parameter 
The weight on the platform is saved as 
a reference when the scale has stabilized.
“Stability” is defined as the point at which
the fluctuation of a measured value lies
within a defined tolerance range. The 
narrower the tolerance range, the more
stable the platform is at stability. 
In the operating menu, under:
appl1: count: 3.11.
you can define whether the value 
is saved when “standard stability” is
reached, or only at “increased stability”
(narrower tolerance range. If you select 
“At increased stability,” the reference 
sample weight stored will be more 
accurate and the results more reproducible,
but the response time of the weighing
instrument might be longer.
Accuracy Level for Calculating 
Reference Weight
The resolution applied for calculating 
the reference weight is defined in the
operating menu under:
appl1: count: 3.9.
The resolution for calculating the 
reference sample weight is increased if 
“+1 decimal place”, “+2 decimal places” 
or “With internal resolution” is selected.
With the “+1 decimal place” setting, the
net value is determined to one additional
decimal place (i.e., display accuracy +10);
the “+2 decimal places” increases display
accuracy +100, and so on up to the 
maximum resolution available. 
Minimum Load
The minimum load required for 
initialization of the weighing platform
is configured in the operating menu
under:
appl1: count: 3.6.
Once the limit is exceeded by the load,
initialization can begin. If the load 
is too light, the following will occur
when you try to save a value:
– The error code 
Inf 29 is displayed
– The weighing platform is not initialized
– The preset reference sample quantity 
is saved
The minimum load required for auto-
matic taring of the container weight on
the platform (“autotare first weight”) 
is configured in the operating menu
under:
Appl: m.weigh: 3.5.
You can choose from the following 
10 levels for this setting:
1 digit
2 digits
5 digits
10 digits
20 digits
50 digits
100 digits
200 digits
500 digits
1000 digits
The “digits” here refer to the scale
intervals in the connected weighing
platform. If the interval of the 
connected weighing platform is 1 g, 
for example, and 1000 digits are
required, you must place at least 
1000 g (=1000 intervals =1000 digits)
on the weighing platform for initiali-
zation.
Application 1: Counting Z