Data Transfer and Statistical Evaluation in the
Automatic Mode
Storing Measured Values
The first value stored must be greater than 30 scale
intervals of the weighing instrument’s display
(e.g. > 3.0 g on a decigram scale). A value can be stored
only when the stability symbol is displayed (i.e., when
fluctuation < 1 scale interval). The first value stored is
used as the reference value for the statistics. The next
value cannot be stored until the weighing instrument
has been unloaded (i.e., until the load on the pan is less
than 30% of the reference value). The next value is
stored only if it is > 70%
of the reference value.
Printer Configuration
Configure the following in the statistics application
menu:
– Statistics program (-A3-) = 1
– Automatic data transfer = 1 (-P6-).
Weighing Instrument Configuration
The following must be configured in the weighing
instrument (see weighing instrument instructions
for details):
– Automatic printing
– Data output without stability
15