Page 4.12
SCHILLER AT-2 6-Channel ECG Unit
SERVICE HANDBOOK Issue 2.1 November 1998
Chapter 4
Module Removal and Replacement
Main Board MK 12-1
Prerequisite
The Warnings and Cautions at the beginning of the chapter must be observed.
The top assembly must be removed and all external cable assemblies disconnected.
The printer and printer tray assembly must be removed
Tools and Equipment
Posi-drive screwdriver
Parts
Main board MK 12-1. Part number as detailed in Chapter 6.
Board Removal
WARNING
ENSURE THAT THE MAINS CABLE IS DISCONNECTED BEFORE COMMENCING
1. Disconnect the following connectors:
live and the neutral bayonet connectors to the mains connector
battery connector
2. Unscrew the 12 spacer fixations (four on the mains transformer) and remove the board.
Board Replacement
To replace the Main board MK 12-1 proceed as follows:
1. Position the board and secure at the 12 fixing points ( four on the mains transformer)
2. Connect the following:
live and the neutral bayonet connectors to the mains connector
live and the neutral bayonet connectors to the battery
3. Replace the paper tray and printer assembly and reassemble the unit.