Page 1.27
Chapter 1
Operating Elements
SCHILLER AT-2plus 6-Channel ECG Unit
SERVICE HANDBOOK Issue d May 2002
Memory and Data Transmission Option (cont.)
Line Transmission
To transmit directly over line do the following
Set Communication mode to LINE - key sequence:
ALT - 0 - 9 - 2 - 1
Connect the cable assembly (optional accessory, art. No. 2.310 159) between the RS-232
connector on the AT-2plus and the COM interface of your Computer.
Ensure that the SEMA communication program (SEMACOMM) is active on the computer
(see SEMA handbook).
Enter the Memory menue by pressing the ALT key followed by the key `M` :
ALT - M
Press and hold the ALT key and then press T.
ALT + T
The message `TRANSMITTING` appears while the unit is sending in the message box
(under the date and time box)
If a transmission error occurs the message `Tx ERROR` is displayed.
Check all settings in the SEMACOMM program (baud rate; parity - none; stop bit - 2; time
between blocks, records - 100ms).
Check that the transmission speed is the same in both the AT-2plus and the SEMACOMM
program.
To stop transmission press and hold the ALT key and then press key `Q`.
ALT + Q