SCHOTT Lab 850 - Unstable Readings and Conditions

SCHOTT Lab 850
73 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
What to do if... Lab 850
56
ba75890e01 11/2009
No stable measured
value
CalClock flashes
Display,
LoBat
– Incorrect buffer solutions – Change calibration
procedure
– Buffer solutions too old – Use only once.
Note the shelf life
– Buffer solutions depleted – Change solutions
Cause Remedy
Cause Remedy
pH electrode:
– Junction contaminated – Clean junction
– Membrane contaminated – Clean membrane
Test sample:
– pH value not stable – Measure with air excluded if
necessary
– Temperature not stable – Adjust temperature if
necessary
Electrode + test sample:
– Conductivity too low – Use a suitable electrode
– Temperature too high
– Organic liquids
Cause Remedy
– Calibration interval expired – Recalibrate the measuring
system
Cause Remedy
– Batteries almost empty – Replace the batteries (see
page 53)