Lab 850 What to do if...
57
ba75890e01 11/2009
Obviously incorrect
measured values
Instrument does not
react to keystroke
You want to know which
software
version is in the
instrument
Cause Remedy
pH electrode:
pH electrode unsuitable Use a suitable electrode
Temperature difference
between buffer and test sample
too high
Adjust temperature of buffer
or sample solutions
Measurement procedure not
suitable
Follow special procedure
Cause Remedy
Operating condition undefined
or EMC load unallowed
Processor reset:
Press and hold the <SC>
key and switch the meter on
Cause Remedy
E. g., a question by the service
department
Switch on the meter.
During the display test,
display the software version
with <OK>.