Activate acoustic
signals (bEEP)
You can set whether an acoustic signal is to be emitted
whenever a key is pressed or a stable weight value has
been attained. The latter is relevant for the Hold/
Autohold function.
1. Select the item "bEEP" from the menu.
2. Confirm the selection.
3. Select a menu item
Press: acoustic signal whenever a key is
pressed
–Hold: acoustic signal with a stable weight
value.
4. Confirm the selection.
The current setting is displayed.
5. Select the setting you require:
–On
–Off
6. Confirm the selection.
You will leave the menu automatically.
7. If you also wish to activate the acoustic signals for
the second function, repeat this process.
Select attenuation (Fil) You can use attenuation (Fil = filter) to reduce any
interference during weighing. The selected setting
affects the sensitivity with which the weight display
responds to patient movements and the time until the
"Hold" function continuously displays a weight value.
1. Select the item "Fil" from the menu.
2. Confirm the selection.
The current setting is displayed.
3. Select an attenuation level.
NOTE
•With the setting "0", it is possible that despite
the "Hold" function being activated, a
continuous weight value is not displayed for
patients who are not so steady.
Fil Weight display Hold
0Sensitive Slow
1Average Average
2Sluggish Fast