Questions and Answers

  • M
    Michelle RuizSep 12, 2025
    Why are some analysis modules not shown in the Analysis tab of Seca Measuring Instruments?
    • C
      Caitlin HernandezSep 12, 2025
      If some analysis modules are not shown in the Analysis tab of your Seca Measuring Instruments, it might be because the list of analysis modules which can be displayed has been restricted by the administrator.
  • M
    melissamillerAug 25, 2025
    What to do if the touchscreen display remains off on Seca Measuring Instruments?
    • D
      Daniel WagnerAug 25, 2025
      If the touchscreen display on your Seca Measuring Instruments remains off, consider these possibilities: * The device might be on standby. Touch the touchscreen display or press the ON/OFF button. * The device may not be switched on. Ensure the device is powered on. * There might be no power supply. Check the power connection.
  • A
    Aaron ChapmanAug 26, 2025
    Why is the charge status LED red on Seca Measuring Instruments?
    • K
      Kevin KrauseAug 27, 2025
      A red charge status LED on your Seca Measuring Instruments indicates: * The rechargeable battery is discharged. Suspend the measuring mat in the storage compartment of the monitor to charge the rechargeable battery. * The rechargeable battery is faulty and needs replacement. Replace the measuring mat. * The inductive charging interface of the monitor is covered. Suspend the measuring mat in the magnetic catch of the monitor, then stow other measuring accessories.
  • D
    Daniel JenkinsAug 27, 2025
    Why is the WiFi LED red on Seca Measuring Instruments?
    • S
      sarahsanchezAug 28, 2025
      A red WiFi LED on your Seca Measuring Instruments indicates: * There is no WiFi connection to the monitor. Check the settings for the WiFi connection and correct if necessary. * The WiFi module of the measuring mat is faulty. Replace the measuring mat.
  • N
    Nicole JonesAug 29, 2025
    Why is the BIA tab not active on Seca Measuring Instruments?
    • M
      Martin ColemanAug 29, 2025
      The BIA tab might not be active on your Seca Measuring Instruments due to: * No seca patient file prepared. Prepare a seca patient file. * Measuring mat not activated. * No measuring mat present.
  • B
    bakerbrandonAug 31, 2025
    Why is the measuring mat not switching on for Seca Measuring Instruments?
    • J
      Jason WatkinsAug 31, 2025
      If the measuring mat for your Seca Measuring Instruments cannot be switched on: * The rechargeable battery might be discharged. Suspend the measuring mat in the storage compartment of the monitor and charge the rechargeable battery. * The measuring mat might be faulty. Replace the measuring mat.
  • K
    Kristen WallaceNov 22, 2025
    Why are the results of bioimpedance measurement deviating significantly from expected results in Seca Measuring Instruments?
    • M
      Michael BurnettNov 22, 2025
      Deviations in bioimpedance measurement results from Seca Measuring Instruments can occur due to: * Mismatch between patient position set on the device and actual patient position. Ensure these match. * Patient movement during the measurement. Ask the patient to remain still. * Patient's arms or legs not held away from the body. Ensure the patient holds their arms and legs away from their body. * Incorrectly assigned electrode cables. Verify that the electrode cables are connected to the correct electrodes according to the patient's position. * Defective electrodes. Replace the electrodes. * Faulty electrode cables or measuring mat. Replace the measuring mat. * Incorrect seca patient file called up. Transmit the measurement to seca analytics 115 PC software and assign it to t...
  • K
    Kara CunninghamSep 13, 2025
    What does it mean when one or more LEDs on the measuring mat are not on for Seca Measuring Instruments?
    • H
      harrisdonnaSep 13, 2025
      If one or more LEDs on the measuring mat are not on for your Seca Measuring Instruments, the measuring mat is likely faulty and needs to be replaced.
  • C
    Cynthia MitchellSep 10, 2025
    Why is there insufficient cuff pressure in Seca Measuring Instruments?
    • L
      Laura CaseySep 10, 2025
      Insufficient cuff pressure in Seca Measuring Instruments can be due to: * Incorrect blood pressure cuff used. Use the correct size of blood pressure cuff and only use seca blood pressure cuffs. * Blood pressure cuff or compressed air tube leaking. Dispose of the blood pressure cuff and use a replacement. * Pump in the device faulty. Do not continue using the device and have it repaired by seca Service.
  • S
    Susan DixonSep 7, 2025
    Why is the password not accepted on Seca Measuring Instruments?
    • K
      Kimberly JonesSep 7, 2025
      If your password is not being accepted on Seca Measuring Instruments, it could be due to: * The password was overwritten on the last synchronization with the seca analytics 115 PC software. Use the new password. If you don't know the new password, contact the administrator. * Following a backup, the old passwords are valid again. Use the old password.