Troubleshooting • 111
English
17-10-05-353-002g_10-2020 B
11.3 Bioimpedance measurement
One or more LEDs on the
measuring mat not on
Measuring mat faulty Replace measuring mat
seca patient files cannot
be transmitted to the
measuring mat
No WiFi connection
Suspend the measuring mat in the
storage compartment of the device
Call up seca patient file again, data will be
transmitted via the infrared interface
Infrared interface faulty Inform seca Service
No WiFi connection
Distance between measuring mat and
monitor is too big
Reduce distance
Performing a measurement (without
WiFi direct connection to the measuring
mat)
Fault Cause Remedy
Fault Cause Remedy
Bioimpedance
measurement not available
BIA measuring mat deactivated
Administrator:
Activate measuring mat: Further
information is available here: Retrofitting
the measuring mat (secamVSA 535 only)
Device is connected to an EMR system
via the
seca connect 103 software
Not a malfunction.
Bioimpedance measurement not possible
when connected via
seca connect 103
Results of bioimpedance
measurement deviate
significantly from expected
results
Patient position set on device does not
match actual patient position
Ensure that the patient position set on the
device and the actual position match
Patient moved during the
measurement
Request the patient not to move during the
measurement and repeat the
measurement
The patient's arms or legs are not held
away from the body
Ask the patient to hold their arms and legs
away from the body
Electrode cables incorrectly assigned
Ensure that the electrode cables are
connected to the correct electrodes to suit
patient position
Electrodes defective Replace electrodes
Electrode cables or measuring mat
faulty
Replace measuring mat
Incorrect seca patient file called up
Transmit measurement to
seca analytics 115 PC software and assign
to the correct seca patient file there
Some analysis modules
are not shown in the
Analysis tab
List of analysis modules which can be
displayed has been restricted by
administrator
Administrator: Modify the analysis modules
which can be displayed
Value of an analysis
parameter is shown in red
Value outside the normal range
determined for that analysis parameter
Repeat the measurement to rule out
measurement error
If the value remains outside the normal
range on a repeat measurement, take the
measured value into account as the
examination continues