Troubleshooting • 115
English
17-10-05-353-002g_10-2020 B
Manual temperature input
SpO
2
measurement
Fault Cause Remedy
Manual temperature input
not possible
Measured values outside permitted
range
Observe permitted temperature range,
Technical data: Manual temperature
input
Device has integrated temperature
measurement (probe or in-ear
thermometer)
Not a malfunction
Use the measuring equipment for the
device
Temperature module of the
SmartBucket activated
Administrator:
Deactivate temperature module:
System\SmartBucket tab
Manual temperature
entries do not appear in
the
seca analytics 115 PC
software
Manual temperature entries are not
synchronized with the
seca analytics 115 PC software
Not a malfunction
Enter values in the
seca analytics 115 PC
software manually
Fault Cause Remedy
Implausible measuring
results
Intra-aortic balloon pump affecting
pulse rate
Check pulse rate using ECG
Sensor damp
Dry sensor
Use dry sensor
Sensor not applied correctly
Apply sensor correctly, see instructions for
use for sensor
Strong ambient light Cover application site with opaque material
Electromagnetic interference
Switch off devices in proximity, isolate
interfering device
Align interfering device differently or set up
in a different location
Increase the distance between this device
and the interfering device
Poor perfusion
Assess patient vital signs using alternative
means
Apply blood pressure cuff to location with
better perfusion
Measurement not possible
Masimo SET
®
only:
Sensor service life expired
Use new Masimo SET
®
sensor.
Sensor or patient cable defective
Dispose of sensor or patient cable, use
spare part
Measurement not possible Sensor or patient cable defective
Dispose of sensor or patient cable, use
spare part
Pulse is not found or is lost
Sensor too rigid
Use suitable size of sensor
Apply sensor to a different finger
Strong ambient light Cover application site with opaque material
Poor perfusion
Assess patient vital signs using alternative
means
Apply sensor to a location with better
perfusion