Technical data • 121
English
17-10-05-353-002g_10-2020 B
12.2 Monitor interfaces and network ports
Interfaces and network ports
Recommended WiFi settings
This table contains settings for optional WiFi performance.
Power consumption
Standby (touchscreen display off, ON/OFF button green)
Operation (ON/OFF button white)
Operation (charging rechargeable battery for monitor and measuring
mat, on/off button is white)
< 5 W
< 9 W
< 35 W
Medical device in accordance with Directive 93/42/EEC Class IIa
IEC 60601-1:
Insulated device, protection class: I
Type of protection
seca mBCA 525 IP20
Type of protection
secamVSA 535 IP21
Duty cycle Continuous duty
Interfaces
seca 360° wireless
2 x USB 2.0 (max 500 mA)
LAN: Ethernet (10/100 Base-T)
WiFi: 2.4 GHz (WPA, WPA2 PSK,
WPA2 Enterprise PEAP RADIUS)
Infrared
Inductive charging of measuring mat
rechargeable battery
Compatible printer
Microsoft
®
Windows
®
-compatible printer
via
seca analytics 115 PC software
Further technical data (all models)
Interfaces Protocol Data transmission rate Factory setting
WiFi: IEEE 802.11 b/g/n Up to 72.2 Mbps Off
LAN: IEEE 802.3u 100 Mbit/s On
TCP port: Transmission Control Protocol n/a 20010, Off
UDP port: User Datagram Protocol n/a 20011
File Transfer port: File Transfer Protocol n/a 20012
seca connect 103 port: Proprietary n/a 22020
USB (2 ports): USB 2.0
480 Mbit/s
per port
On
seca 360° wireless: Proprietary 50 kb/s Off
Settings Recommended settings Consequences in the event of different settings
Authentication/
encryption
WPA2 Personal
WPA2 Enterprise (EAP-TLS)
No network connection, other encryption methods
not supported
Frequency band
Single band 2.4 MHz
IEEE 802.11 b/g/n
Data transmission rate Up to 72.2 Mbps
Network configuration
DHCP
UDP
Firewall/
ports to be opened
TCP port: 20010
UDP port: 20011
File transfer port: 20012
seca connect 103 port:
22020
No synchronization
Separate VLAN No special requirements
QoS No special requirements
VoiP No special requirements
WiFi multimedia No special requirements
Network latency No special requirements
IT support No special requirements