50 • secamVSA 535/seca mBCA 525
17-10-05-353-002g_10-2020 B
Measuring blood pressure
WARNING!
Hazard to patient
Do not use luer lock adapter on the blood pressure measurement
tubes. The use of luer lock adapters can lead to accidental connec-
tion of the blood pressure tubes to intravenous lines and thus lead
to an infusion of air into the bloodstream of the patient.
The decision to use this device with pregnant or preclamptic
patients is the responsibility of the user.
The device has no alarm function. Never leave the patient
unobserved during a measurement.
Frequent measurements may lead to perfusion disorders and
consequently to severe harm to the patient.
Route the compressed air tube so that it cannot kink. A kinked com-
pressed air tube leads to sustained cuff pressure. This may lead to
perfusion disorders and consequently to severe harm to the patient.
Do not place the blood pressure cuff over open wounds. This may
lead to further harm to the patient.
Do not apply any external pressure onto the blood pressure
measurement tubes or onto the blood pressure cuff.
For patients with moderate to severe cardiac arrhythmias, inaccu-
rate measurements may result with the blood pressure measure-
ment.
The following factors may affect the measurement result:
Measuring location (e.g. altitude)
Patient position (standing, sitting, lying)
Physiological state of the patient (e.g. exertion, movement,
shaking, shivering)
Patient's age
Arteriosclerosis
Poor perfusion
Diabetes
Kidney disease
If the measurements seem implausible, check and evaluate the vital
signs of the patient using alternative means. Then check the mea-
surement function of the device with the help of the section
"Troubleshooting".
WARNING!
Patient hazard, incorrect measurement
Do not place the blood pressure cuff on the arm on the side on
which a mastectomy has been performed.
Apply the blood pressure cuffs so that the blood circulation of the
patient is not compromised.
Do not apply the blood pressure cuffs on spots that have weak
circulation or on extremities on which intravenous routes lie.
High cuff pressures may be unpleasant for sensitive patients. Keep
the patient’s general condition under observation during the
measurement.
Preparing the blood pressure measurement
Starting a single measurement
Starting a multiple measurement
Modifying presets