170 • secamVSA 535/seca mBCA 525
17-10-05-353-002g_10-2020 B
5.2 Setting up a WiFi connection (mobile operation)
Introduction For mobile use, for example on the ward of a hospital, you can integrate the
device in your network as a WiFi client to exchange data with the
seca analytics 115 PC software or seca connect 103.
A requirement for exchanging data with the
seca analytics 115 software is
that the
seca analytics 115 PC software is installed on a server or standalone
PC and access data for your WiFi network are known.
A requirement for exchanging data with the
seca connect 103 software is
that the
seca connect 103 software is installed on a server and the configura-
tion of the
seca connect 103 software (server IP and connect port) is known.
For devices with bioimpedance measurement: If you integrate the device in
your network via WiFi, the measuring mat can only communicate with the
monitor via the infrared interface.
NOTICE!
Malfunction
The measuring mat and the monitor can communicate via the infrared
interface only when the measuring mat is suspended in the magnetic
catch of the monitor.
► Ensure that the measuring mat is suspended in the magnetic catch
of the monitor before and after every measurement.
➔ Introduction
➔ Activating the WiFi connection
➔ Deactivating the WiFi connection
Server/Stand-Alone-PC
WiFi-Router
Infrared
WiFi WiFi
seca mBCA 525/
seca mVSA 535
Monitor
seca 531
Measuring mat