Troubleshooting • 119
English
17-10-05-353-002g_10-2020 B
11.8 seca 360° proximity connection
11.9 seca 360° wireless connection
Fault Cause Remedy
Weight or height values do
not appear in the display
No
seca 360° proximity connection set
up
Administrator:
Setting up a seca 360° proximity
connection
Network connection interrupted
LAN: Ensure that the network cable is cor-
rectly plugged in and is not damaged
WiFi: Reduce distance from router
seca scale/measuring rod incorrectly
configured
Administrator: Configure seca scale/
measuring rod as follows:
Scan user ID: Required
Scan patient ID: Required
Confirmation of measurement on seca
measuring device: Required
Port used: Port configured for communi-
cation with the
seca connect 103 software
(default: 22020) Setting up a connec-
tion to the seca analytics 115 PC software
Follow seca 103/452 system instructions
for use
No measurement has yet been
performed using the seca scale/
measuring rod
Perform measurement as described in the
instructions for use for the seca scale/seca
measuring rod
seca scale/measuring rod has not
submitted a value
Submit value as described in the
instructions for use for the seca scale/seca
measuring rod
An error message appears
in the Weight or Height
field
An error occurred on the seca scale/
measuring rod
Eliminate error as described in the instruc-
tions for use for the seca scale/seca
measuring rod.
Follow seca 103/452 system instructions
for use
If you are unable to eliminate the error,
contact seca Service.
Fault Cause Remedy
Weight or height values do
not appear in the display
seca 360° wireless wireless module of
the connected scale/measuring rod not
activated
Administrator:
Setting up a seca 360° wireless
connection
No
seca 360° wireless wireless
connection set up
Scale/measuring rod not wireless-
capable (do not contain a
seca 360° wireless wireless module)
Enter measured values manually:
Preparing a bioimpedance
measurement