124 • secamVSA 535/seca mBCA 525
17-10-05-353-002g_10-2020 B
12.5 Vital signs measurement
Blood pressure measurement
Blood pressure measurement
COVIDIEN
TM
FILAC
TM
3000 temperature measurement
COVIDIEN
TM
GENIUS
®
3 temperature measurement
Manual temperature input
Masimo SET
®
SpO
2
measurement
seca SpO
2
measurement
seca blood pressure module
Measuring method
Oscillometric
Possible to switch between inflation
and deflation
Single measurement
Multiple measurement can be config-
ured: Max. 6 measurements in max.
30 minutes
Initial cuff pressure, deflation
Maximum cuff pressure
Can be set: 80 mmHg – 280 mmHg
(260 mmHg cannot be set)
300 mmHg
Measuring range for blood pressure
Inflation:
Systolic BP
Diastolic BP
Mean arterial pressure
Deflation:
Systolic BP
Diastolic BP
Mean arterial pressure
77 mmHg – 200 mmHg
45 mmHg – 190 mmHg
56 mmHg – 193 mmHg
25 mmHg – 280 mmHg
10 mmHg – 220 mmHg
15 mmHg – 260 mmHg
Accuracy (under laboratory conditions, verified with CuffLink patient
simulator from Fluke)
max. ± 3 mmHg / 2 %,
the larger value in each case applies
Measuring accuracy for blood pressure (determined by the manufacturer
of the measuring module in clinical trial in accordance with
DIN EN ISO 81060)
Inflation:
Mean deviation of systolic BP
Standard deviation of systolic BP
Mean deviation of diastolic BP
Standard deviation of diastolic BP
Deflation:
Mean deviation of systolic BP
Standard deviation of systolic BP
Mean deviation of diastolic BP
Standard deviation of diastolic BP
0.36 mmHg
4.27 mmHg
-0.12 mmHg
3.78 mmHg
0.10 mmHg
3.24 mmHg
-0.20 mmHg
2.95 mmHg
Pressure transducer:
Accuracy
Resolution
±1 mmHg
1 mmHg
Leakage rate
Limit value for pressure
Switch off and release pressure at (first fault)
< 3 mmHg/min
300 mmHg
> 330 mmHg
Measuring time for blood pressure:
Normal
Maximum (adults)
15 – 20 s
90 s