Technical data • 125
English
17-10-05-353-002g_10-2020 B
COVIDIEN
TM
FILAC
TM
3000
temperature measurement
Pulse rate:
Measuring range for inflation
Measuring range for deflation
Accuracy (under laboratory conditions, verified with CuffLink patient
simulator from Fluke)
45 min
-1
– 200 min
-1
30 min
-1
– 240 min
-1
max. ± 3 min
-1
/ 3 %,
the larger value in each case applies
IEC 60601-1: Medical electrical device, type BF (defibrillation-protected)
seca blood pressure module
COVIDIEN
TM
FILAC
TM
3000 temperature module
Measurement modes Direct, predictive
Measuring position:
Blue probes
Red probes
Oral, axillary
Rectal
Measuring range
Direct mode
Predictive mode
30 °C – 43 °C (86 °F – 109.4 °F)
35.5 °C – 42 °C (95.9 °F 107.6 °F)
Measuring time (following application at measuring position)
Direct:
All measuring positions
Predictive:
Oral, no fever
Oral, fever
Axillary
Rectal
60 – 120 sec
3 – 5 sec
8 – 10 sec
8 – 12 sec
10 – 14 sec
Switchover time from predictive mode to direct mode
Measuring position not detected (following removal from probe cover)
No stable temperature value obtained (following application)
60 sec
70 sec
Accuracy (water bath):
Direct mode
Predictive mode
± 0.1 °C ( ± 0.2 °F)
± 0.1 °C (± 0.2 °F)
IEC 60601-1: Medical electrical device, type BF
Type of protection against ingress of liquids IP21
Clinical accuracy
a
b
Measuring position: Oral Axillary Rectal
d (age group I)
-0.44 -0.01 0.09
L
A
(age group I)
1.01 0.86 0.99
d (age group II)
-0.21 -0.04 0.12
L
A
(age group II)
0.75 0.65 0.67
σ
r
0.34 0.28 0.28
a.The clinical accuracy of the COVIDIEN
TM
FILAC
TM
3000 was determined in a clinical study in accordance with EN 80601-2-
56. Clinical bias d
and limits of agreement L
A
are quoted for the age group and measuring position in question. Clinical re-
peatability σ
r
is independent of age. The reference body sites of the reference thermometer used in the clinical study corre-
spond to the measuring positions quoted.
b.The age of the test subjects in age group I is between 3 and 4 years. The age of the subjects in age group II is 5 and above.