102 • secamVSA 535/seca mBCA 525
17-10-05-353-002g_10-2020 B
8. HYGIENE TREATMENT
WARNING!
Electric shock
The device is not de-energized when the on/off key is pressed and the
display goes out. Use of fluids on the device may cause electric
shock.
► Ensure that the device is switched off before performing any hygiene
treatment.
► Disconnect the power supply connector before performing any
hygiene treatment.
► Before each hygiene treatment, take the rechargeable battery out of
the device (if present and removable).
► Ensure that no fluids penetrate the device.
NOTICE!
Damage to device
Inappropriate detergents and disinfectants may damage the sensitive
surfaces of the device.
► Use only disinfectants free of chlorine and alcohol which are explic-
itly suitable for acrylic sheet and other sensitive surfaces (active in-
gredient: quaternary ammonium compounds, for example).
► Do not use caustic or abrasive detergents.
► Do not use organic solvents (e.g. white spirit or petroleum spirit).
► Use disinfectants including the active ingredient 70 % isopropanol
only for measuring accessories for measuring vital signs.
8.1 Cleaning
► Clean the device and its accessories as described in the table:
➔ Cleaning ➔ Removing/fitting probe holder (devices
with temperature probe)
➔ Disinfecting ➔ Removing/fitting the magazine holder
(devices with in-ear thermometer)
➔ Sterilizing
Component (depending on version) Interval Cleaning
seca mBCA 525: Monitor with storage
compartment
secamVSA 535: Monitor with
SmartBucket
As required
1.Remove all the measuring accessories (measuring
devices and consumables) from the device (depending
on version)
• ➔ Removing the probe holder
• ➔ Removing the magazine holder
2.Moisten a soft cloth with a soap solution
3.Wipe over all surfaces
4.Allow to air-dry for approx. 30 minutes
Measuring mat and electrode cables As required
1.Moisten a soft cloth with a mild soap solution
2.Clean measuring mat and electrode cables
3.Allow to air-dry for approx. 30 minutes
Blood pressure cuff and compressed air
tube
As required
1.Moisten a soft cloth with a mild soap solution
2.Clean blood pressure cuff and compressed air tube
3.Rinse thoroughly with water
4.Allow to dry at room temperature