102 • secamVSA 535/seca mBCA 525
17-10-05-353-002g_10-2020 B
8. HYGIENE TREATMENT
WARNING!
Electric shock
The device is not de-energized when the on/off key is pressed and the
display goes out. Use of fluids on the device may cause electric
shock.
Ensure that the device is switched off before performing any hygiene
treatment.
Disconnect the power supply connector before performing any
hygiene treatment.
Before each hygiene treatment, take the rechargeable battery out of
the device (if present and removable).
Ensure that no fluids penetrate the device.
NOTICE!
Damage to device
Inappropriate detergents and disinfectants may damage the sensitive
surfaces of the device.
Use only disinfectants free of chlorine and alcohol which are explic-
itly suitable for acrylic sheet and other sensitive surfaces (active in-
gredient: quaternary ammonium compounds, for example).
Do not use caustic or abrasive detergents.
Do not use organic solvents (e.g. white spirit or petroleum spirit).
Use disinfectants including the active ingredient 70 % isopropanol
only for measuring accessories for measuring vital signs.
8.1 Cleaning
Clean the device and its accessories as described in the table:
Cleaning Removing/fitting probe holder (devices
with temperature probe)
Disinfecting Removing/fitting the magazine holder
(devices with in-ear thermometer)
Sterilizing
Component (depending on version) Interval Cleaning
seca mBCA 525: Monitor with storage
compartment
secamVSA 535: Monitor with
SmartBucket
As required
1.Remove all the measuring accessories (measuring
devices and consumables) from the device (depending
on version)
Removing the probe holder
Removing the magazine holder
2.Moisten a soft cloth with a soap solution
3.Wipe over all surfaces
4.Allow to air-dry for approx. 30 minutes
Measuring mat and electrode cables As required
1.Moisten a soft cloth with a mild soap solution
2.Clean measuring mat and electrode cables
3.Allow to air-dry for approx. 30 minutes
Blood pressure cuff and compressed air
tube
As required
1.Moisten a soft cloth with a mild soap solution
2.Clean blood pressure cuff and compressed air tube
3.Rinse thoroughly with water
4.Allow to dry at room temperature