Hygiene treatment • 103
English
17-10-05-353-002g_10-2020 B
Temperature probe (red/blue) with cable As required
1.Discard probe cover and dispose of it
2.Moisten a soft cloth with a mild soap solution
3.Clean temperature probe
4.Shake out temperature probe so that no liquid remains
in it
5.Allow to air-dry for approx. 30 minutes
Probe holder (red/blue) As required
1.➔ Removing the probe holder
2.Moisten a cotton bud with a mild soap solution
3.Wipe over all the surfaces of the probe holder
In-ear thermometer
After every
use
In-ear thermometer and cable:
1.Do not discard probe cover
2.Moisten a soft cloth with a mild soap solution: Ratio of
water to soap solution: 20:1, temperature: Max. 55 °C
(130 °F)
3.Wring out the cloth so that no excess liquid can
penetrate the in-ear thermometer
4.Wipe over in-ear thermometer and cable
5.Dry the in-ear thermometer and cable using a lint-free
cloth
6.Discard probe cover
As required
Measuring head and lens:
1.Carefully remove all foreign particles using a cloth
moistened with alcohol (70 % isopropanol)
2.Wipe dry the lens at the tip of the measuring head with
a lint-free cloth (e.g. spectacle-cleaning cloth)
3.Make sure that the lens at the tip of the measuring
head does not have any fingerprints or marks on it
4.Allow the in-ear thermometer to dry completely in air
Magazine holder for probe covers (in-ear
thermometer)
As required
1. ➔ Removing the magazine holder
2.Moisten a soft cloth or cotton bud with a mild soap
solution
3.Wipe over the surfaces of the magazine holder
4.Allow the magazine holder to dry completely in air
SpO
2
sensor with cable
Follow the manufacturer’s instructions for use
Patient cable for SpO
2
sensor
Follow the manufacturer’s instructions for use
Component (depending on version) Interval Cleaning