154 • secamVSA 535/seca mBCA 525
17-10-05-353-002g_10-2020 B
3. Press the SpO
2
button.
The presets are displayed (in this case: Masimo SET
®
pulse oximetry):
4. Press the desired sensitivity:
NOTE
The Maximum setting (Masimo SET
®
pulse oximetry) is not available
as a preset. Select this setting for each measurement directly
➔ Measuring oxygen saturation (SpO
2
).
5. Press the Confirm button.
The dialog window closes.
The modified settings are adopted.
Presets for temperature
(COVIDIEN
TM
FILAC
TM
3000 only)
1. Press the Settings tab.
Masimo SET
®
SpO
2
module
Mode Indication
Normal
• Normal perfusion
• Mild perfusion disorders
Adaptive Probe Off Detection
(APOD)
Vigorous patient movements
Maximum
• Poor perfusion
• Severely disrupted signal, for
example due to indoor lighting or
direct sunlight
seca SpO
2
module
Mode Motion tolerance
Stable High
Normal Normal
Sensitive Low