72 • secamVSA 535/seca mBCA 525
17-10-05-353-002g_10-2020 B
Modifying presets
NOTE
Your settings apply to the current measurement procedure. If you
stop the measurement procedure Stopping a measurement proce-
dure, the presets defined by the administrator become active again.
1. Ensure that no SpO
2
sensor has been applied to the patient.
2. Press the SpO
2
field.
The SpO
2
settings dialog window opens.
The presets are displayed (in this case: Masimo SET
®
pulse oximetry).
3. Press the desired sensitivity:
4. Press the Confirm button.
The dialog window closes.
The changed settings are used for the current measurement.
5. Start the SpO
2
measurement as described in the section entitled
Starting an SpO
2
measurement.
Weight and height The device provides several options for adding the parameters weight and
height. Your administrator specifies which of the options you can use. Further
information is available here: Adding weight and height.
Masimo SET
®
SpO
2
module
Mode Indication
Normal
Normal perfusion
Mild perfusion disorders
Adaptive Probe Off Detection
(APOD)
Vigorous patient movements
Maximum
Poor perfusion
Severely disrupted signal, for
example due to indoor lighting or
direct sunlight
seca SpO
2
module
Mode Motion tolerance
Stable High
Normal Normal
Sensitive Low