Modifying presets
NOTE
Your settings apply to the current measurement procedure. If you
stop the measurement procedure ➔ Stopping a measurement proce-
dure, the presets defined by the administrator become active again.
1. Ensure that no SpO
2
sensor has been applied to the patient.
2. Press the SpO
2
field.
The SpO
2
settings dialog window opens.
The presets are displayed (in this case: Masimo SET
®
pulse oximetry).
3. Press the desired sensitivity:
4. Press the Confirm button.
The dialog window closes.
The changed settings are used for the current measurement.
5. Start the SpO
2
measurement as described in the section entitled
➔ Starting an SpO
2
measurement.
Weight and height The device provides several options for adding the parameters weight and
height. Your administrator specifies which of the options you can use. Further
information is available here: ➔ Adding weight and height.
Masimo SET
®
SpO
2
module
Mode Indication
Normal
• Normal perfusion
• Mild perfusion disorders
Adaptive Probe Off Detection
(APOD)
Vigorous patient movements
Maximum
• Poor perfusion
• Severely disrupted signal, for
example due to indoor lighting or
direct sunlight
seca SpO
2
module
Mode Motion tolerance
Stable High
Normal Normal
Sensitive Low