112 • secamVSA 535/seca mBCA 525
17-10-05-353-002g_10-2020 B
11.4 Vital signs measurement
General
Blood pressure measurement
General
Blood pressure measurement
COVIDIEN
TM
FILAC
TM
3000 temperature measurement
COVIDIEN
TM
GENIUS
®
3 temperature measurement
Manual temperature input
SpO
2
measurement
Fault Cause Remedy
Vital signs tab is not
displayed
USB connecting cable of the
SmartBucket not connected to the
monitor
Connect the USB connecting cable of the
SmartBucket to the monitor
Restart the device
SmartBucket not activated
Administrator:
Activate SmartBucket: Managing
system components in the administrator
menu
secamVSA 535 is being
operated on a wheeled
stand: Measuring results
for vital signs are
implausible
Original seca USB connecting cable
not being used
USB extension cable also in use
Use original seca USB connecting cable
(in scope of delivery for the wheeled stand)
Do not use a USB extension cable
Fault Cause Remedy
Implausible measuring
results
Excessive patient movement
Ask the patient to move as little as possi-
ble
Incorrect blood pressure cuff used
Use correct size of blood pressure cuff
Only use seca blood pressure cuffs
Blood pressure cuff not applied
correctly
Correctly apply blood pressure cuff, see
instructions for use for blood pressure cuff
Blood pressure cuff applied to an
extremity to which there is intravenous
access
Apply blood pressure cuff to a different
extremity
Insufficient cuff pressure
Incorrect blood pressure cuff used
Use correct size of blood pressure cuff
Only use seca blood pressure cuffs
Blood pressure cuff or compressed air
tube leaking
Dispose of blood pressure cuff, use
replacement
Pump in device faulty
Do not continue using device and have it
repaired by seca Service
No option to select
Multiple measurement in
the NIBP settings dialog
window
Administrator has configured a
measuring profile. The name of the
measuring profile is displayed.
Not a malfunction
Press the name of the measuring profile
Starting a multiple measurement.
Blood pressure values do
not appear in the EMR
system or in the
seca analytics 115 PC
software
Multiple measurement was performed
For current measurement:
Modifying presets and Starting a
single measurement
Administrator:
Specify Individual measurement as preset
Presets for blood pressure