Troubleshooting • 113
English
17-10-05-353-002g_10-2020 B
COVIDIEN
TM
FILAC
TM
3000
temperature measurement
COVIDIEN
TM
GENIUS
®
3 temperature
measurement
Fault Cause Remedy
No temperature
measurement possible
Temperature module of the
SmartBucket not activated
Administrator:
Activate temperature module:
System\SmartBucket tab
Implausible measuring
results,
temperature measurement
unsuccessful
Special patient condition, such as
hypothermia
Assess patient vital signs using alternative
means
Switch from predictive measurement
mode to direct mode
Rectal measurement performed with
blue temperature probe
Only perform rectal measurement with red
temperature probe
Oral/axillary measurement performed
with red temperature probe
Only perform oral/axillary measurement
with blue temperature probe
Patient activity before oral temperature
measurement:
Physical exertion
Eating/drinking
Brushing teeth
Smoking
Only perform oral temperature
measurement about 20 minutes after any
of these activities
Set measuring position does not match
actual measuring position
Select measuring position to suit the
probe used
Set correct measuring position on device
No probe cover used
Disinfect the temperature probe
Disinfecting
Use probe cover
Temperature probe faulty
Dispose of temperature probe, use
replacement
Temperature probe cannot
be pushed completely into
the probe holder
Probe cover not discarded
Carefully withdraw temperature probe and
probe cover from the probe holder
Discard probe cover
Push the temperature probe into the
probe cover
Fault Cause Remedy
No temperature
measurement possible
Temperature module of the
SmartBucket not activated
Administrator:
Activate temperature module:
System\SmartBucket tab
Patient temperature exceeds
measuring range of in-ear thermometer
Assess patient vital signs using alternative
means
Patient temperature undershoots
measuring range of in-ear thermometer