7.3 Resetting the user interface
Use the Reset GUI function to reset the user interface (GUI = Graphical User
Interface) of the device to factory settings ➔ Overview of factory settings.
seca patient files and user accounts will be retained in this process.
NOTE
If all seca patient files and user accounts are to be removed from the
device, proceed as described in the section entitled ➔ Resetting the
device.
1. Press the Reset tab.
2. Press the Reset GUI button.
The user interface will be reset to factory settings ➔ Overview of factory
settings.
seca patient files and user accounts will be retained.
7.4 Exporting the system log/audit trail
This function allows you to export the system log/the audit trail and make it
available to seca Service for support purposes, for example.
1. Press the System tab.
2. Press the Data management button.
3. Press the Export button.
4. Select the export destination:
► USB memory stick: continue at step 5.
► Network: Continue at step 6.