116 • secamVSA 535/seca mBCA 525
17-10-05-353-002g_10-2020 B
11.5 Network connection
Fault Cause Remedy
No WiFi connection
WiFi function of device is deactivated
Administrator:
Activate WiFi
Distance between monitor and WiFi
router is too big
• Reduce distance
• Transmit measuring results to
seca analytics 115 PC software or
seca connect 103 via LAN
WiFi not available in your institution
Transmit measuring results to
seca analytics 115 PC software or
seca connect 103 via LAN
No network connection
Firewall: ports required are not enabled
Administrator:
Enable required ports in firewall ➔ Monitor
interfaces and network ports
Firewall/gateway configuration does
not permit use of LAN and WiFi in
parallel
Administrator:
Deactivate one of the two transmission
options on the device.