122 • secamVSA 535/seca mBCA 525
17-10-05-353-002g_10-2020 B
12.3 Measuring mat
12.4 Bioimpedance measurement
Measuring method
Redundant
power supply
No special requirements
Settings Recommended settings Consequences in the event of different settings
Measuring mat
Dimensions
Depth
Width
Height
783 mm
170 mm
20 mm
Net weight approx. 1 kg
Ambient conditions, operation
Temperature
Air pressure
Humidity
+10 °C to +40 °C (50 °F to 104 °F)
700 hPa 1060 hPa
20 % – 80 %, non-condensing
Ambient conditions, storage
Temperature
Air pressure
Humidity
-10 °C to +60 °C (14 °F to 140 °F)
700 hPa 1060 hPa
15 % – 95 %, non-condensing
Ambient conditions, transport
Temperature
Air pressure
Humidity
-10 °C to +60 °C (14 °F to 140 °F)
700 hPa 1060 hPa
0 % – 95 %, non-condensing
Setup location, maximum altitude above mean sea level 3000 m
Power supply
Range (measurements)
Lithium-ion battery
approx. 5 h
Medical device in accordance with Directive 93/42/EEC Class IIa
FDA classification Class II Medical Device
IEC 60601-1: Medical electrical device, type BF
Type of protection IP44
Duty cycle Continuous duty
Measuring method
Clinical studies
Accuracy of predictive formulas
Measuring method
Measuring method
8-point bioimpedance measurement
4-point bioimpedance measurement
(measure right half of body)
Measuring frequencies 1; 2; 5; 10; 20; 50; 100; 200; 500 kHz
Measured values
Impedance (Z), resistance (R),
reactance (X
c
), phase angle (φ)
Phase angle measuring range 0° to 20°
Impedance measuring range 10 Ω to 1000 Ω
Measuring segments
Right arm, left arm, right leg, left leg,
right half of body, torso
Measuring current 100 μA (+20 %, -50 %)
Measuring time max. 30 s