Operation • 53
English
17-10-05-353-002g_10-2020 B
If inflation fails to deliver a measured value, the device automatically
switches to deflation. If necessary, the device re-pumps several
times during the deflation (re-pumping: Increasing cuff pressure by
approx. 50 mmHg followed by incremental releasing of that pres-
sure). If no measured value is delivered even after several re-pump-
ing steps, the procedure is canceled.
Starting a multiple measurement
Press the button on the monitor.
The first part-measurement starts (in this case: 6 part-measurements).
Current cuff pressure is displayed.
The symbols for measurement procedure and measuring mode flash (in
this case: Inflation).
The part-measurement stops automatically as soon as valid blood
pressure values are detected.
The values for systolic/diastolic blood pressure SYS/DIA and mean
arterial pressure MAP are displayed.
The waiting time until the next part-measurement starts.
The next part-measurement starts automatically.
Once all part-measurements are present, mean blood pressure (mean
value) is displayed.
Disregarded part-measurements are displayed struck through (in this
case: Measurement 1).
You have the following options for continuing:
Measure more vital signs Measuring vital signs
Measure bioimpedance Measuring bioimpedance
Stop the measurement procedure Stopping a measurement
procedure
NOTE
Use the Start button to cancel the mean measurement and continue
or to cancel and re-start.
If inflation does not deliver a measured value, the device automati-
cally switches to deflation.
If necessary, the device re-pumps several times during the deflation
(re-pumping: Increasing cuff pressure by approx. 50 mmHg fol-
lowed by incremental releasing of that pressure). If no measured
value is delivered even after several re-pumping steps, the proce-
dure is canceled.