Making settings for measuring mode • 155
English
17-10-05-353-002g_10-2020 B
2. Press the Vital signs button.
3. Ensure that the temperature probe is pushed into the probe holder
completely.
4. Press the TEMP field.
The presets are displayed (in this case: COVIDIEN
TM
FILAC
TM
3000 blue):
5. Press the desired measurement method.
Predictive
Direct
6. Press the desired measuring position (COVIDIEN
TM
FILAC
TM
3000 blue
only).
Oral
Axillary
7. Press the Confirm button.
The dialog window closes.
The modified settings are adopted.