94 •
If you find faults or deviations during the function
check, as a first step try to remedy the fault with the aid
of the section entitled "Troubleshooting" from page 94.
CAUTION!
Bodily injury
If you find faults or deviations during the func-
tion check which you are unable to remedy with
the aid of the section entitled "Troubleshooting"
from page 94, you may not use the device.
Have the device repaired by seca service or
by an authorized service partner.
Please see the section entitled "Servicing" on
page 96.
10.TROUBLESHOOTING
Problem Cause / Solution
No weight displayed
with a load on the scale.
The scale has no power supply.
- Check whether the scale is switched on.
- Check whether batteries are inserted.
0.00 does not appear
before weighing.
A load was already on the scale before it was
switched on.
- Remove the load from the scale.
- Switch the scale off and back on again.
One segment is illumi-
nated either continu-
ously or not at all.
The corresponding point has a fault.
- Inform the service department.
The display
appears.
Battery power is running down.
- Insert new batteries as soon as possible.
The display "bAtt"
appears.
Batteries are flat.
- Insert new batteries.
The display "StOP"
appears.
Maximum load has been exceeded.
- Remove the load from the scale.
The display "tEMP"
appears.
The ambient temperature of the scale is too high
or too low.
- Set up the scale in an ambient temperature
between +10 °C and +40 °C.
- Wait for around 15 minutes until the device
has adapted to the ambient temperature.