11.SERVICING
On leaving the factory, your seca scale has an accu-
racy of ±0.15 % or better. To preserve this level of
accuracy, the product must be set up with care and
serviced regularly. Depending on how frequently the
scale is used, we recommended servicing at intervals
of 3 to 5 years.
ATTENTION!
Incorrect measurements as a result of poor
servicing
Have servicing and repairs carried out
exclusively by seca service or by an
authorized service partner.
You can find service partners in your area at
www.seca.com or by sending an e-mail to
service@seca.com.
The display "Er:X:16"
appears.
The scale was caused to oscillate, the zero point
could not be determined.
- Restart the scale.
The Enter key
(send/print) is pressed
and the display "Er:X:71"
appears.
No data transmission possible, wireless module
deactivated.
- Activate wireless module (see "Activating the
wireless module (SYS)" on page 88).
The Enter key
(send/print) is pressed
and the display
"Er:X:72" appears.
No data transmission possible, no wireless group
set up.
- Set up wireless group (see "Setting up a
wireless group (Lrn)" on page 88).