KemiDose | pH · ORP · Chlorine
0000139024 rev. 1.0 EN
8
7. ORP CALIBRATION
Reference calibration
5 s
Set Redox calibration
60 s
465 mV calibration
H
2
O
9
Save and exit
10
11
H
2
O
H
2
O
5
6
7
3
8
465 mV
4
1
2
The unit will flash a temperature
value
Set the temperature value
measured with the instrument
Ex. 750 mV
CAL Reference
720 mV
CAL Reference
750 mV